Introduction
The Pulse Oximeter adopts Photoelectric Oxyhemoglobin Inspection
Technology combined with Capacity Pulse Scanning & Recording
Technology to measure SpO2 and pulse through the finger, which is
applicable for use in family, hospital, oxygen bar, community
healthcare, sport healthcare (it can be used before or after doing
sports, and it is not recommended to use the device during the
process of having sport) and etc.
Functions
1)Low-voltage indication: the low-voltage indication symbol appears
when the voltage is too low to work.
2)Automatic standby function.
3)Data storage function.
4)The data can be uploaded to terminal equipment by wireless mode.
5)Charging function.
Performance
SpO2 measurement range: 0 % ~ 100 % (resolution: 1 %)
Accuracy: 70 % ~ 100 %, ±2%; 0 %~69 %, unspecified
PR measurement range: 20 bpm ~ 250 bpm (resolution: 1bpm)
Accuracy: ±2 bpm or ±2 %, whichever is greater
Accuracy under conditions of low perfusion: low perfusion 0.4%,
SpO2 accuracy: ±4%; PR accuracy: ±2 bpm or ±2%, whichever is
greater.
Light interference: compared to the values measured in normal
condition and ambient light condition, the deviation is less than
±1 %.
Power consumption: ≤ 100 mA
Working voltage: DC 3.6V ~ 4.2V
Battery: one 3.7 V rechargeable lithium battery
Working time for battery: 8 hours (theoretical value)
Battery lifetime: ≥ 500 times (charge and discharge times)
Safety classification: internally powered equipment, type BF
applied part
Accessories
One SpO2 probe
One User Manual
One USB data line
One Bluetooth adapter (optional)
One power adapter (optional)